Customization: | Available |
---|---|
Theory: | Autoclave |
Type: | Ethylene Oxide Sterilization Cabinet |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Sterilization room material | 304corrosion proof stainless steel |
Outer case material | Carbonsteel (1.5mm thick)baking plastic powder |
Temperature of sterilization | 50ºC+5ºC |
Working pressure | -60Kpa |
The highest pressure value | -140kpa |
Sterilization humidity | 45%-85%RH |
Power supply | AC380V |
Selection of gas | Ethylene oxide gas and carbon dioxide mixed gas |
The dosage of gas | 0.6g/L |
Sterilization time | Adjustable 0-99 bours |
Vacuum speed | 15minutes to vacuum-60kpa |
Air replacement | Automatic ventilation 3times,periodic intervals for30minutes |
Full automatic control system | One button start to the end,the whole process automatically complete paint records |
Display status | Display temperature,humidity,pressure,sterilization time,processing time,the number of residual gas |
Printer | The whole process of sterilization information records to print,in order to prepare for verification |
Humidity sensor | Humidity,pressure and temperature sensors are used to detect the humidity,pressure andtemperature in the chamber automatically,which ensures the data to be kept in a percise range. |
Pressure sensor | |
Temperature sensor | |
Vacuum system | Adopting the oil free.vacuum pump,which is simple and convenient,the vacuum speed is fast,and the service life is long. |
Sealing door | Using electric automatic doors,pneumatic sealing easy operation,good sealing |
Heating system | Explosion proof electric heating device is adopted,which is more convenient and quick. |
Multiple protection function | With over-temperature protection function,more safe and reliable |
Sealant strip | Concave and sealing ring (patent product)silica gel material |
Shelf | 304stainless steel |
Push cart | Carbon steel material |