Ethylene Oxide Sterilizer (large manual door)
Vacuum speed |
25 minutes to -60kpa |
Sterilization room material |
304 anticorrosive stainless steel |
Standard sterilization time (min) |
0-99 hours adjustable |
Sterilization temperature(ºC) |
50ºC±3ºC |
Gas selection |
Pure ethylene oxide gas |
Power connection |
380V/50Hz |
Outer box material |
Carbon steel (thick 1.8mm) baking powder |
Humidity during sterilization |
45%-75% |
Working pressure |
-60kpa |
Ethylene oxide sterilization concentration |
1g/L |
Air replacement:Automatic ventilation, ventilation frequency and interval time can be set freely
Automatic control system:one button start to end, the whole process automatically complete with the print record
Display: state display temperature, humidity and pressure, sterilization time, processing time, the number of residual gas
Printer:the whole process of sterilization information recording and printing, for verification check.
Humidity sensor:built-in humidity sensor automatic detection of cavity humidity, to ensure that the cavity humidity is not less than 45%RH.
Pressure sensor:built-in pressure sensor automatically detects cavity pressure, to ensure the accurate and effective pressure inside the chamber.
Temperature sensor:built-in temperature sensor, automatic detection of cavity temperature, to ensure that the cavity temperature is maintained at 50 degrees C + 3 temperature within the precise range.
Humidifying system:pulse humidification under vacuum condition, ensure the humidity of the cavity is even and controllable; the humidity of the sterilizing room should be controlled within the relative humidity range of 45%-80%;
Residual processing device filter on particle diameter of more than 0.3 m the filtering rate of not less than 99.5%.
Sterilization room: leak rate determination of leakage rate at no-load, sterilization room temperature constant, test pressure -50kPa, time 1H were determined under the conditions of determination of time average leakage rate less than 0.16kPa/min.
Rate:vacuum under no-load condition, atmospheric vacuum sterilizer from time to -60kPa should be less than 30min.
Vacuum system:using oil-free vacuum pump, simple and convenient, fast vacuum speed, long service life.
Heating system adopts explosion proof electric heating device, which is more convenient and quick.
Pallet rack:use 304 stainless steel pallet rack
A variety of protective functions: with over-voltage, over temperature protection function, more safe and reliable.
Casters:Yes
Alarm system:with high temperature alarm system, ultra low pressure alarm system; when sterilization room temperature exceeds 59 degrees, should be able to sound alarm.
Category: class II medical devices